| CEQ 2000 DNA Analysis System |
Eight-capillary array, 100
µm i.d., 50 cm long
Four-wavelength fluorescence detection/capillary
96-well microplate format for samples
96-well microplate format for buffer
Automatic gel replenishment
Automatic sample denaturation and injection
Automatic base calling and reporting
|
Weight:
Height:
Width:
Depth:
Power:
Excitation: |
155 lbs (70.5 kg)
37 in (94 cm)
24 in (61 cm)
24 in (61 cm)
100-240 VAC, 5 amps, 50/60 Hz
Two diode lasers
Class 1 laser hazard |
| |
|
| CEQ 2000 Controller |
Personal computer with the
following recommended specifications:
|
Pentium* II, 350 MHz |
4.2 GB hard disk
- 64MB RAM, 2MB video RAM
- CD-ROM
- CD Writer
- Super VGA 17-in display (1024 x 768)
- Power: 100-240 VAC, 8 amps, 50/60 Hz
|
|
|
|
|
| CEQ 2000 Software |
| Windows**-based CEQ sequencing
software on CD and installed on CEQ workstation
Data file format:
Staden, FASTA, SEQ, Tab-delimited text, PHRED compatible |
| |
|
|
| Dye-Terminator Cycle Sequencing |
Optimized DTCS methods using four colors/samples
500 bases at 98% accuracy***
Eight samples read to 500 bases in two hours; 96/24hours
At least 100 runs per capillary; eight capillaries /array, depending
upon metohd used
100 samples/gel cartridge
100 samples/DTCS kit |
| |
|
|
| Other Genetic Analysis Products From
Beckman Coulter |
| Product |
Biomek®
2000 Automated Laboratory Workstation
MultimekTM
96 96-Channel Automated Pipettor
Molecular
Biology Core System
Fluorescent
Differential Display
Oligo
1000M DNA Synthesis System
Dye-Terminator